Product Numbers:
Product description:AL-B3000 TV Shadowless Operating Lamp(Camera on the handle)
Shadowless operating lamp with battery
● B3000 series shadowless Operating lamp can widely be used in diversified operation occasion to meet the demand of the illumination, and it is the ideal illumination instrument of the modern operating theater.
● The Whole reflection optical system is made by CAD/CAM technology, its illumine depth is up to 1050mm ;by using color temperature compensation technology, the color reversion is uprised, and it is more suitable for the illumination of the complex surgery such as encephalon surgery and thorax surgery.
● Micro computer digital control, with 8 shifts luminance for selection and it offers illuminance memory funcition and wide voltage working speciality, cam work stably under AC 150V~250V, with strong anti-jamming capacity.
● Full close streamline lamp body design completely meets the demand of the sterilization and high standard of laminar flow purification in the operation room.
● The panel is made of German PC (anti-bullet plastic) by injection moulding, makes the light softer and more reliable.
● Focusing system has a delicate structure, and easy to change bulb and the ceramic seat; detachable handle cover can take high temperature (≤134°) sterilization treatment.
● With back up bulb automatic switch function, when the main bulb was burnt, auxiliary lamp will be lighted in 1/2 second. There’s failure indication in the handle control panel, to remind the bulb to be changed after operation.
● Offering soft start up unit, extending the serve life of the bulb effectively. Adopting the imported long life halide-tungsten bulb, the bulb life can be up to 1000 hours.
● All the key parts use the world famous products, ensure stable and reliable operation; flame retardant level material can endure high temperature and safe, meets IEC norm.
● Equipped with imported balance arm suspension system, which is light and handy, six group joints move together and moves smoothly, positions stablely. 360 universal design can satisfy various needed height and angle in the operation.
● Top-bulit (bottom board ) power supply system, avoid troubles in installation of hang-on-wall control box.
☆ Optimized laminar flow purification, sterilization,mirror face design and back-up bulb replacement
● Full closed steamline lamp head which is designed according to aerodynamics is helpful to laminar flow purification and application of sterilization in the operating room, this can reduce the risk of operation infection.
● The pervious glass at the bottom of lamp head is made of anti-shock and anti-disinfectant material. This pervious glass is durable and easy cleaning, and won’t be effected by radiation , cleanance and long time use.
● The sterilization handle can move up and down, revolve the lamp head and adjust the focus. It can be detached after operation and steeped in the liquid or by high temperature high pressure sterilization
● The operating lamp adopts automatic monitor and back-up bulb automatic replacement system which can supply the illuminance during the whole operation, it won’t affect the operation. When change the bulb, you just need to unscrew the three screws of the handle body and detach it, it is easy to replace the bulb. The handle body is equiped with convenient and sterilizing handle.
● Multi-lens reflecting system is designed by computer professional technology, there are more than 3800 pieces multi-lens reflecting system which makes the light beam focus as a high illuminance symmetrical light cylinder.
☆ Natural operation color temperature
The CRI for natural light is 100. The CRI of B series integral reflecting shadowless operating lamp is 98,which is very near natural light. It can help the surgeon to distinguish the color of operating region(including blood vessel and nerve tract),it can help to distinguish the tissues of body clearly.
☆ Advanced digital integrated circuit control panel
● It has functions of swith, intensity adjustment and back-up bulb test; Also has functions of 8-shift intensity selection and intensity memory.LED intensity indicative bar can show the needed intensity in the operation intuitively.
● There is a main bulb fault indicator on the control panel which remind the staff to replace the bulb after the operation.
● It has characteristic of wide power working, which can work stably between AC180V-250V with high anti-interference ability.
☆ Safe backup illuminance
When the main bulb is broken down in the operation process,the backup bulb will
be lighted withi 0.2 second,the
intensityand illuminance center
won’t be effected. The circuit
board is seted by a soft start
system which extend the bulb lifetime,
it is better for long time operation illuminance.
☆Camera Technical Data:
Definition: | ≥600TVL |
SNR: | ≥46db |
Shoot distance(m) | Main Vidicon:0.8-2 Panorama vidicon:1-10 |
Power supply voltage | AC 200V,50Hz-60Hz/100V,60Hz |
Foci | F3.66-91.36mm |
Zooming | 22double |
Video frequency export | 1.0Vp-p75Ω |
Vidicon power supply | 12VDC±10% |
Video frequency transceiver | plast-off frequency 1.2GHz |
Lowest height of installation(mm) | 3100 |
Cinerama camera
☆ Technical Data:
Illuminance(Lux): | 80,000-150,000Lux 60,000-120,000Lux |
Color temperature: | 4000±500 |
Spot diameter(mm): | 160-280 |
Illuminance depth(mm): | ≥1000 |
Lamp head diameter: | 600mm |
Brightness adjustment: | 8-stage continuous light adjustment |
Total irradiance: | B3000 700:426W/㎡ B3000 500:580W/㎡ |
The ratio between irradiance and illuminance: | B3000 700: ≦3.3mW/㎡lx B3000 500: ≦4.7mW/㎡lx |
Color reduction index(Ra): | B3000 700: 94 B3000 500: 93 |
Power supply voltage: | AC 200V/50-60Hz |
Input power(W): | 400 |
24V Bulb power(W): | 150/150 |
Shift of major and stand by lamps: | Automatic |
Lowest herght of installation(mm) | 2900 |
☆ Installation Diagram: